TABLE 2

Contrast, Background Variability, and Average Lung Residual for 8:1 Sphere-to-Background Ratio on Vision, Directly Compared with Published Results from mCT Flow

Contrast (%)Background variability (%)
Sphere size (mm)VisionmCT FlowVisionmCT Flow
1086.841.96.06.3
1377.263.15.05.4
1785.068.13.94.4
2289.876.63.33.6
2887.471.33.03.0
3789.677.72.22.4
Average lung residual (%)3.512.1
  • mCT Flow measurements include effect of low-resolution matrix and postreconstruction 3-mm gaussian filter (12).