TABLE 4

Image Quality Results for 4:1 and 8:1 Sphere-to-Background Ratios for Vereos PET/CT System (NEMA NU2-2012)

Contrast recoveryBackground variability
Sphere diameter (mm)4:18:14:18:1
1054.3 (48.6–61.5)54.4 (53.2–55.4)8.8 (7.7–10.2)9.3 (8.9–9.8)
1375.3 (68.6–78.8)75.9 (74.3–79.0)7.2 (6.2–8.3)7.5 (7.1–7.9)
1780.4 (78.5–83.1)81.6 (80.0–83.4)5.7 (5.0–6.3)5.7 (5.2–6.1)
2283.9 (80.7–88.3)86.5 (86.2–87.1)4.4 (4.1–4.7)4.3 (3.9–4.8)
2881.4 (08.2–82.8)82.5 (82.1–83.3)3.3 (3.1–3.6)3.5 (3.2–3.9)
3787.0 (86.8–87.3)85.8 (86.1–85.4)2.5 (2.5–2.3)2.6 (2.1–2.9)
Lung residual6.356.36
  • Data are mean percentages followed by ranges in parentheses.