TABLE 1

Spatial Resolution Measured for PET Component of Vereos PET/CT System (NEMA NU2-2012)

Vereos
ParameterDistance (cm)FWHM (mm)FWTM (mm)Discovery MI FWHM (mm) (8)mCT Flow FWHM (mm) (20)
Transverse14.24* ± 0.078.40* ± 0.154.15*4.33*
Axial14.17 ± 0.178.80 ± 0.214.484.25
Transverse radial104.55 ± 0.098.87 ± 0.135.475.16
Transverse tangential104.35 ± 0.018.97 ± 0.044.494.72
Axial104.39 ± 0.009.08 ± 0.026.015.85
Transverse radial205.84 ± 0.0610.43 ± 0.047.535.55
Transverse tangential204.92 ± 0.019.96 ± 0.124.906.48
Axial204.60 ± 0.139.21 ± 0.266.107.80
  • * Average value of radial and tangential measurements.

  • FWTM = full width at tenth maximum.

  • Data are average ± SD of the measurements at the axial center an off center position. Reported spatial resolutions of Discovery MI (SiPM-based system with 20-cm axial FOV (8)) and mCT Flow (PMT-based system with 22.1-cm axial FOV (20)) are shown for comparison. Spatial resolution measurements for 15-cm axial FOV Discovery MI and 16.2-cm axial FOV mCT Flow are not available but are expected to be similar to those of extended-axial-FOV systems (25,35).