TABLE 1

Percentage Error for Various ROIs for 18F-FMISO Validation Study Comparing PET-Based and 4D CT-Based Motion Estimation

ROI regionCTAC-basedNAC-basedAIM
Liver-dome13.9% ± 19.9%4.9% ± 8.7%3.2% ± 4.4%
Large tumor (mid right)2.2% ± 7.0%1.8% ± 6.3%−1.3% ± 2.0%
Small tumor (mid left)−7.1% ± 2.6%−6.3% ± 2.5%−4.3% ± 2.2%
  • Data are μErr% (g) ± SD. SD is calculated across all gates.