Scanning and Reconstruction Specifications
Scanner | Symbia T16 SPECT/CT | Precedence SPECT/CT | Biograph TrueV 40 PET/CT | Biograph mCT PET/CT | Biograph mMR PET/MRI |
Emission data | |||||
Views | 180 | 128 | |||
Time per view (s) | 6 | 8 | |||
Time per bed position (min) | 2 | 2 | 5 | ||
Axial field of view per bed position (cm) | 38.7 | 38.1 | 21.8 | 21.8 | 25.8 |
Matrix | 128 × 128 | 128 × 128 | 336 × 336 | 400 × 400 | 344 × 344 |
Slice thickness (mm) | 3 | 2 | 2 | ||
Energy window (keV) | 140% ± 15% | 140% ± 15% | 435–650 | 435–650 | 430–610 |
Reconstruction method | 5 iterations | 4 iterations | 3 iterations | 3 iterations | 3 iterations |
10 subsets | 16 subsets | 21 subsets | 21 subsets | 21 subsets | |
Astonish | 4 mm | 4 mm | 4 mm | ||
CT acquisition | |||||
Voltage (kV) | 130 | 140 | 120 | 120 | |
Tube current | Weight adapted (dynamic) | Weight adapted (dynamic) | 40 mAs (dynamic) | 40 mAs (dynamic) | |
Rotation (s/rotation) | 0.6 | 0.5 | 0.5 | 0.5 | |
Pitch | 1.5 | 0.938 | 0.8 | 0.8 | |
Slice thickness (mm) | 5 | 5 | 3 | 2 |