TABLE 1

Scanning and Reconstruction Specifications

ScannerSymbia T16 SPECT/CTPrecedence SPECT/CTBiograph TrueV 40 PET/CTBiograph mCT PET/CTBiograph mMR PET/MRI
Emission data
 Views180128
 Time per view (s)68
 Time per bed position (min)225
 Axial field of view per bed position (cm)38.738.121.821.825.8
 Matrix128 × 128128 × 128336 × 336400 × 400344 × 344
 Slice thickness (mm)322
 Energy window (keV)140% ± 15%140% ± 15%435–650435–650430–610
 Reconstruction method5 iterations4 iterations3 iterations3 iterations3 iterations
10 subsets16 subsets21 subsets21 subsets21 subsets
Astonish4 mm4 mm4 mm
CT acquisition
 Voltage (kV)130140120120
 Tube currentWeight adapted (dynamic)Weight adapted (dynamic)40 mAs (dynamic)40 mAs (dynamic)
 Rotation (s/rotation)0.60.50.50.5
 Pitch1.50.9380.80.8
 Slice thickness (mm)5532