TABLE 6

Diagnostic Accuracy

Consensus readingOptimistic analysisPessimistic analysis
ModalitySensSpecPPVNPVACCSensSpecPPVNPVACCSensSpecPPVNPVACC
pBS64.396.575.094.391.956.397.075.093.391.568.881.236.794.379.5
SPECT/CT60.092.856.393.888.456.393.156.393.188.0*62.589.147.693.885.5
PET/CT80.097.985.796.995.575.098.086.796.194.9*81.393.165.096.991.5
PET/MRI85.794.975.097.493.585.794.975.097.493.585.794.975.997.493.5
  • Sens = sensitivity; Spec = specificity; NPV = negative predictive value; ACC = accuracy.

  • PET/CT showed significantly higher overall accuracy than SPECT/CT with optimistic analysis (*P = 0.039). PET/CT showed significantly higher specificity (P = 0.012) and overall accuracy (P = 0.011) than pBS with pessimistic analysis. All other comparisons were nonsignificant (P > 0.05).