TABLE 1

PET Uptake Differences Between Reference PET/CTAC and PET/MRAC (ZTACUC, ZTACSEC, and ATAC) Across All 240 VOIs

Difference
PET/MRACkBq/mL%P
PET/ZTACUC0.79 ± 0.525.6 ± 3.5<0.01
PET/ZTACSEC0.03 ± 0.330.2 ± 2.40.23
PET/ATAC−0.10 ± 0.71−0.9 ± 5.00.03
  • Differences are mean ± SD. A paired t test was performed for pair of PET/CTAC and PET/MRAC.