TABLE 1
Uniform Phantom Results for Static Brain Protocol During T0 Period
Manufacturer
No. of scanners
Volume average SUV
*
MAD
*
GE
36
1.00 ± 0.03
2.9% ± 1.2%
Philips
7
0.99 ± 0.03
8.5% ± 0.7%
Siemens
21
0.98 ± 0.05
3.4% ± 1.4%
Combined
64
0.99 ± 0.04
3.7% ± 2.1%
↵
*
Reported as mean ± SD.