TABLE 4

Image Quality Using NEMA NU2-2012 for Different Reconstruction Algorithms and LBR of 8:1

Contrast recovery (%)Background variability (%)
Sphere diameter (mm)VPHDVPHD-SQ.ClearVPHDVPHD-SQ.Clear
1031 ± 230 ± 233 ± 35.5 ± 0.84.8 ± 0.63.7 ± 0.1
1344 ± 149 ± 151 ± 15.0 ± 0.64.4 ± 0.53.4 ± 0.1
1757 ± 267 ± 270 ± 14.5 ± 0.33.9 ± 0.33.1 ± 0.2
2267 ± 170 ± 177.5 ± 0.43.9 ± 0.23.5 ± 0.22.9 ± 0.2
2860 ± 360 ± 268 ± 23.4 ± 0.23.4 ± 0.12.7 ± 0.1
3763 ± 262 ± 277.3 ± 13.5 ± 0.33.8 ± 0.13.0 ± 0.1
Lung residual (%)25.5 ± 0.326.5 ± 0.411.5 ± 0.5
  • Data are mean ± SD. Matrix was 256 × 256. Q.Clear used a β of 350. VPHD and VPHD-S used 4 iterations, 12 subsets, and 4.8-mm gaussian filter.