TABLE 1

Mean Absolute Percentage Error in SUVmax and Percentage Error Reduction Due to Inclusion of TOF Information, for Each Clinically Relevant Location

Mean absolute percentage error in SUVmax ± SD
LocationImplantTransaxial/axial diameter of ellipsoid signal void (mm)Non-TOF (%)TOF (%)Error reduction due to TOF (%)
JawDental implants23.4/13.91.0 ± 1.00.7 ± 0.926.6
Humeral headShoulder prostheses60.9/36.11.3 ± 2.21.2 ± 2.37.5
ChestInjection systems23.4/13.90.5 ± 0.50.2 ± 0.359.8
SternumMetal wires23.4/13.94.3 ± 4.5*2.2 ± 2.9*49.2
Thoracic and lumbar spineSpinal fusion23.4/13.90.3 ± 0.30.3 ± 0.3−2.3
Femoral headHip prostheses60.9/36.117.9 ± 25.113.5 ± 13.524.7
  • * Statistically significant difference, P = 0.02.