%diff and |%diff| in 18F-FDG Uptake (kBq/mL) Between Atlas-/ZTE-AC and CT-AC in Each Merged Region
Parameter | Frontal lobe | Occipital lobe | Parietal lobe | Insula and cingulate | Central structure | Temporal lobe | Cerebellum |
%diff on atlas-AC | −1.17% ± 2.61% | −1.64% ± 1.41% | −1.20% ± 1.76% | −1.10% ± 2.31% | −1.83% ± 2.21% | −2.67% ± 2.07% | −3.64% ± 2.65% |
%diff on ZTE-AC | −1.57% ± 1.25% | 0.05% ± 0.89%* | −1.59% ± 1.10% | −0.55% ± 1.17% | 0.54% ± 1.35%* | 2.46% ± 1.19%* | 3.31% ± 1.70%* |
|%diff| on atlas-AC | 2.42% ± 1.35% | 1.66% ± 1.39% | 1.63% ± 1.32% | 2.22% ± 1.10% | 2.54% ± 1.20% | 2.78% ± 1.90% | 3.73% ± 2.51% |
|%diff| on ZTE-AC | 1.75% ± 0.95% | 0.64% ± 0.58% | 1.70% ± 0.89% | 1.06% ± 0.67%* | 1.03% ± 0.99%† | 2.46% ± 1.19% | 3.31% ± 1.70% |
%diff is atlas-AC/ZTE-AC minus CT-AC divided by CT-AC. |%diff| is absolute value of %diff.
atlas-AC vs. ZTE-AC, *P < 0.05, †P < 0.01.