Reference | Country | n | Age (SD) | % male | Type of AC | Tracer | Test interval | CAD definition | AC sensitivity | AC specificity | NAC sensitivity | NAC specificity |
Yamauchi (7) | Japan | 150 | 70 (7) | 68 | SSPAC | 99mTc | 3 mo | >50% | 0.91 | 0.90 | 0.77 | 0.76 |
Patil (8) | United States | 54 | 66 (11) | 74 | 153Gd | 99mTc | 30 d | >70% | 0.89 | 0.75 | ||
LM > 50% | ||||||||||||
Sharma (18) | India | 171 | 55 (10) | 82 | CT AC | 99mTc | 3 mo | >50% | 0.57 | 0.89 | 0.65 | 0.83 |
Genovesi (29) | Italy | 104 | 64 (10) | 79 | CT AC | 99mTc | 1 mo | >70% | 0.75 | 0.81 | ||
LM > 50% | ||||||||||||
Huang (30) | China | 99 | 62 (12) | 56 | CT AC | 99mTc | 60 d | >70% | 0.92 | 0.79 | 0.95 | 0.63 |
Bateman (9) | United States | 92 | 64 (12) | 60 | 153Gd | 99mTc | 181 d | >70% | 0.83 | 0.71 | 0.77 | 0.67 |
Wolak (10) | United States | 114 | 65 (21) | 0 | 153Gd | 99mTc | 60 d | >70% | 0.81 | 0.73 | 0.80 | 0.73 |
Taneja (31) | United Kingdom | 141 | 67 (-)* | 40 | 153Gd | 99mTc | 3 mo | >70% | 0.71 | 0.91 | 0.61 | 0.83 |
Masood (19) | United States | 118 | 61 (12) | 67 | CT AC | 99mTc | 3 mo | >50% | 0.94 | 0.59 | 0.93 | 0.56 |
Utsunomiya (32) | Japan | 30 | 68 (-)* | 60 | CT AC | 201Tl | 30 d | >50% | 0.76 | 0.93 | 0.67 | 0.86 |
Thompson (33) | United States | 116 | 60 (11) | 70 | 153Gd | 99mTc | 60 d | >70% | 0.86 | 0.79 | 0.89 | 0.50 |
Banzo (34) | Spain | 99 | 59 (-)* | 72 | 153Gd | 99mTc | 3 mo | >70% | 0.76 | 0.71 | 0.92 | 0.46 |
Links (35) | United States | 66 | (-)* | 68 | 153Gd | 99mTc | Trial period | >50% | 0.96 | 0.85 | 0.78 | 0.62 |
99mTc | 201Tl | Trial period | ||||||||||
Lee (20) | Korea | 68 | 59 (12) | 43 | 99mTc | 99mTc | 2 mo | >70% | 0.74 | 0.75 | 0.67 | 0.59 |
Hendel (36) | United States | 112 | 61 (12) | 56 | 153Gd | 99mTc | 3 mo | >50% | 0.78 | 0.44 | 0.76 | 0.50 |
Gallowitsch (37) | Austria | 107 | 64 (10) | 65 | 241Am | 201Tl | 1–14 d | >70% | 0.94 | 0.91 | 0.79 | 0.80 |
Ficaro (17) | United States | 60 | 63 (12) | 63 | 241Am | Dual tracer† | 90 d | >50% | 0.88 | 0.82 | 0.84 | 0.46 |