TABLE 1

Absolute Detection Efficiency and Detection Limit for Microfluidic and Catheter-Based Detectors

Chip 1 (0.66 μL)Chip 2* (1.46 μL)PE10 (1.85 μL)PE50 (7.93 μL)
IsotopeEβmean (keV)β/CE emission probability (%)η (cps/Bq)AminAmax (Bq/μL)η (cps/Bq)Amin – Amax (Bq/μL)η (cps/Bq)AminAmax (Bq/μL)η (cps/Bq)AminAmax (Bq/μL)
64Cu190, 27837.2, 17.820%100–5.0⋅10520%45–2.3⋅1053%254–1.3⋅1062%93–4.6⋅105
18F25096.739%51–2.5⋅10539%23–1.2⋅1057%103–5.1⋅1054%38–1.9⋅105
11C38699.847%42–2.1⋅10547%19–9.6⋅10415%46–2.3⋅10512%14–6.9⋅104
68Ga353, 8361.2, 87.743%47–2.3⋅10543%21–1.1⋅10522%33–1.6⋅10517%10–4.9⋅104
99mTc120, 1388.8, 14%527–2.6⋅1064%238–1.2⋅1060.2%4,447–2.2⋅1070.2%1,034–5.2⋅106
  • * Estimated.

  • Eβmean = mean energy of β-particles.

  • Detection volume is indicated in brackets for each device. Data for microfluidic Chip 2 were computed.