Scanner | Manufacturer | Transaxial | Axial |
Discovery LS | GE Healthcare | 5.2 | 7.5 |
Discovery ST | GE Healthcare | 6.7 | 5.8 |
Allegro | Philips Healthcare | 5.7 | 6.7 |
SET-2400W | Shimadzu | 5.0 | 6.7 |
Biograph 2 | Siemens Healthcare | 7.1 | 7.0 |
Biograph 16 | Siemens Healthcare | 4.8 | 5.5 |
ECAT Exact HR+ | Siemens Healthcare | 5.4 | 5.3 |
Sensation 16 | Siemens Healthcare | 6.8 | 7.1 |
Data are full width at half maximum at 10 cm according to National Electrical Manufacturers Association NU 2-2001 standards as implemented in PVEC approach used in this study.