TABLE 2

Results of Analyses of Spatial Alignment for Both Patient Groups

ParameterStandard CT and PETTriggered CT and PETP
Mismatch of lung–liver boundary (mm)9.2 ± 8.14.5 ± 6.7<0.001*
Average distance between lesion centroids (mm)6.3 ± 4.05.6 ± 4.20.424
Jaccard similarity coefficient0.30 ± 0.210.32 ± 0.200.609
SUVmax (g/cm3)10.5 ± 6.710.9 ± 6.7<0.001*
SUVmean (g/cm3)6.1 ± 4.06.4 ± 4.00.001*
  • * Significant difference between 2 groups.

  • Numbers are averages and SD for all patients (i.e., scanned on Biograph 40 mCT and on Biograph 64 mCT Flow scanner).