TABLE 3

Recommended Adult Reference Levels for 99mTc-MDP Bone SPECT and 18F-FDG Whole-Body PET

Radiation protection: injected activity
NCRP 172IAC surveyImage quality: ADP*
ExaminationSNMMIADDRLADDRLADDRL
99mTc-MDP bone SPECT740–1,110 MBq (20–37 mCi)833 MBq (23 mCi)1,185 MBq (32 mCi)925 MBq (25 mCi)999 MBq (27 mCi)23,100 MBq⋅min (625 mCi⋅min)25,000 MBq⋅min (675 mCi⋅min)
18F-FDG whole-body PET148–740 MBq (4–20 mCi)555 MBq (15 mCi)710 MBq (19 mCi)518 MBq (14 mCi)592 MBq (16 mCi)1,554 MBq⋅min (42 mCi⋅min)1,776 MBq⋅min (48 mCi⋅min)
  • * 99mTc-MDP bone SPECT values assume use of dual-head system with 25-min study duration. 18F-FDG whole-body PET values assume acquisition with 3 min per bed position.