Subgroup Analyses of Lesion SUV Parameters Based on Size Criteria
SUV parameters | GeminiTF | DigitalTF | P | Change (%) | P |
Lesion SUVmax | 0.078* | ||||
Lesions < 1.5 cm | 4.6 (1.2–19.7) | 6.3 (0.9–24.1) | <0.001† | 37 | |
Lesions ≥ 1.5 cm | 10.0 (3.0–19.3) | 11.1 (3.2–22.4) | <0.001† | 11 | |
Lesion–to–blood-pool SUVmax ratio | 0.02* | ||||
Lesions < 1.5 cm | 2.3 (0.7–5.9) | 4.0 (0.6–9.3) | <0.001† | 72 | |
Lesions ≥ 1.5 cm | 7.1 (2.0–15.8) | 8.7 (2.7–19.5) | <0.001† | 23 | |
Lesion SUV gradient | 0.107* | ||||
Lesions < 1.5 cm | 1.8 (0.3–10.1) | 2.6 (0.6–14.2) | <0.001† | 44 | |
Lesions ≥ 1.5 cm | 3.1 (0.8–9.3) | 3.7 (0.7–10.7) | 0.002† | 19 |