TABLE 3

Summary of the reaction conditions for the radiosynthesis of [18F]FLT on EWOD chip based on Method 2.

ProcessTemperature (°C)Reaction time (min)ReagentConcentration (mM)Droplet volume (μL)Fluorination efficiency (%)*Radioactivity remaining on chip (%)Crude radiochemical yield (%)*
1—Load 18F-TBAFRoom temperatureTBAHCO3456Not applicableNot applicable
2—MeCN azeotropic distillation1052MeCNNot applicable9Not applicableNot applicable
3—Fluorination120318F-FLT precursor90291 ± 4 (n = 10)94 (n = 4)Not applicable
4—Hydrolysis951.5HCl757Not applicable85 (n = 4)82 ± 10 (n = 11)
  • * Fluorination efficiencies and yields are reported with SD based on number of experiments, n.

  • To determine radioactivity remaining on chip, chip was removed after each step and radioactivity on entire chip was measured using dose calibrator for radioactivity measurement.