TABLE 3

Sample Protocol for Single-Day Rest–Stress 99mTc-Labeled Scans with SPECT MPI

FeatureRest studyStress studyTechnique
Dose296–444 MBq (8–12 mCi*)890–1,330 MBq (24–36 mCi*)Standard
PositionSupineSupineStandard
ProneProneOptional
Upright/semiuprightUpright/semiuprightOptional
Delay time (intervals)
 Injection to imaging30–60 min15–60 minStandard
 Rest to stress30 min to 4 hStandard
Energy window15%–20% symmetricSameStandard
CollimatorLow energy high resolutionSamePreferred
Orbit180° (45° RAO to 45° LPO)SamePreferred
Orbit typeCircularSameStandard
NoncircularSameStandard
Pixel size
 64 by 64 matrix6.4 ± 0.4 mmSameStandard
 128 by 128 matrix8–9 mmSameOptional
Acquisition typeStep and shootSameStandard
ContinuousSameOptional
Number of projections60–64SameStandard
Matrix60 × 64SameStandard
Time/projection25 s20 sStandard
ECG gatedOptionalStandardPreferred
Frames/cycle88Standard
1616Optional
R-to-R window (for ejection fraction)100%100%Preferred
Attenuation correctionMeasured attenuation correction before or after SPECT scanMeasured attenuation correction before or after SPECT scanOptional
  • * Lower stress and rest doses may be used with new scanners with solid-state detectors and new software reconstruction.

  • 64 projections are preferable for 201Tl.

  • RAO = right anterior oblique; LPO = left posterior oblique.

  • Adapted and reproduced with permission from (5).