Test | Release criterion |
Appearance of final product | Clear with no visible impurities |
Filter test | >45 psi |
pH | 3.0–5.5 |
Radiochemical purity | >90% (achieved >95%) |
Endotoxin level | <17.5 Endotoxin units per mL |
Radionuclidic purity | Within ±90 s of 68Ga half-life |