TABLE 1
Simulated Gating Error Results in Significant Drop-Off in PSD
Severity
Gating error
Change in PSD
P
vs. baseline
1
0.1% ± 0.1%
−11.8% ± 7.7%
0.01
2
0.8% ± 0.2%
−23.5% ± 9.4%
0.001
3
2.4% ± 0.4%
−41.0% ± 12.2%
<0.001
4
6.3% ± 0.3%
−57.2% ± 11.8%
<0.001
5
12.9% ± 0.3%
−69.2% ± 9.2%
<0.001