TABLE 2

Technical Parameters for Different MRI Sequences

RegionSequenceImage planeSlice thickness (mm)Gap (%)Slices (no.)Acquisition time (min:s)TR/TEMatrixField of view (mm)Resolution (mm2)GRAPPABreath holdRemarks
AC
Whole body3D T1 VIBE DixonCoronal3.1201280:193.60/1.23–2.46*192 × 1215004.1 × 2.62+MRI-based AC
Diagnostic sequences for whole-body coverage
Whole bodyT1 TSECoronal53025–361:11600/8.7384 × 2304502.0 × 1.22+ for chest, abdomen
Whole bodyT2 STIRCoronal53025–361:25–3:157.9/500259 × 3844502.9 × 1.83Free breathing
Diagnostic sequences for partial-body coverage
Partial bodyT2 HASTE fat satAxial520251:081,600/95320 × 2603801.5 × 1.22Free breathingPartial body = chest to pelvis, dark-blood pulse
Lungs3D VIBE + CMAxial40720:183.24/1.23320 × 2404001.7 × 1.32+11% slice oversampling
Abdomen to pelvis3D VIBE + CMAxial50440:163.29/1.16320 × 2604601.8 × 1.42+Two stacks; 45% slice oversampling
Prostate cancer recurrence
ProstateT2 TSEAxial320314:474,610/101320 × 3102000.6 × 0.62
ProstateDWIAxial3.60205:394,500/93160 × 1202602.2 × 1.62b = 50, 400, 800 s/mm2
Prostate3D TWISTAxial3.60205:244.83/1.87192 × 1322602.0 × 1.42DCE; temporal resolution, 4.25 s
Primary staging of ENT tumors
NeckT1 TSE ± CMAxial4.020303:461,330/10384 × 3842300.6 × 0.62Each sequence before and after CM
NeckT2 STIRAxial4.020303:485,810/54320 × 3202300.7 × 0.72
NeckT1 TSE fat sat + CMCoronal4.020203:371,290/9.6320 × 3202300.7 × 0.72
NeckT2 STIRSagittal4.02284:115,090/57384 × 3452400.7 × 0.62
Focus on potential liver metastases
LiverT2 HASTECoronal5.020300:421,400/96256 × 2563801.5 × 1.53+
LiverT2 HASTE fat satAxial5.020351:081,600/95320 × 2603801.5 × 1.22+
LiverDWIAxial5.020355:047,800/82192 × 1443802.6 × 2.02Free breathingb = 50, 400, 800 s/mm2
LiverVIBEAxial3.020640:184.06/1.91320 × 2403801.6 × 1.22+DCE: without CM, arterial, portal-venous, late phase; 13% slice oversampling
Diagnostic whole-body protocol
BrainT2 FLAIRAxial4.030252:089,000/85256 × 1922201.1 × 0.92
BrainT1 GRE +CMAxial4.030252:10250/2.48320 × 2562200.9 × 0.7
BrainT1 GRE +CMCoronal4.030351:25250/2.48320 × 2562200.9 × 0.7
LungsT2 HASTECoronal8.030300:22650/28320 × 2564001.6 × 1.33+
LiverDWIAxial5.020354:089,200/85192 × 1443802.6 × 2.02Free breathingb = 50, 800 s/mm2
NeckVIBE DixonAxial4.020360:285.61/2.46–3.69320 × 2604001.5 × 1.3244% slice oversampling
  • * Fat saturation techniques with Dixon require 2 repetition times.

  • Acquisition time is dependent on number of averages and slices per slab specific for different body regions (3–5).

  • Only sequences not part of other protocols are mentioned.

  • 3D = 3-dimensional; b = b-value; BLADE = noncartesian k-space data acquisition for motion correction in TSE sequences; CM = contrast medium; DCE = dynamic contrast enhanced; DWI = diffusion-weighted imaging; FLAIR = fluid-attenuated inversion recovery; GRAPPA = generalized autocalibrating partially-parallel acquisition (as integrated parallel-acquisition technique [iPAT; Siemens Healthcare]); GRE = gradient echo; sat = saturation; STIR = short-τ inversion recovery; TE = echo time; TR = repetition time; TSE = turbo spin echo; TWIST = dynamic contrast-enhanced sequence with k-space sharing.