TABLE 2

Spatial Resolution Results

FWHM/FWTM (mm) at axial center of FOVFWHM/FWTM (mm) at ¼ axial offset
System energy window reconstructionRadial offset (mm)RadialTangentialAxialRadialTangentialAxial
microPET P4, 350–650 keV, Fourier rebinning + 2D FBP52.29/4.032.18/3.812.20/4.522.34/4.222.14/3.771.75/4.22
102.41/4.232.23/3.922.38/4.662.37/4.142.22/3.841.97/4.49
152.42/4.192.28/3.832.42/4.682.39/4.162.27/3.872.04/4.53
252.61/4.672.25/3.762.42/4.672.53/4.412.30/3.912.07/4.50
503.27/6.402.40/4.102.58/5.093.20/6.082.45/4.292.30/4.74
753.92/8.072.64/4.532.88/5.993.78/7.122.81/5.152.72/5.58
microPET R4, 350–650 keV, Fourier rebinning + 2D FBP52.13/4.902.21/4.222.72/5.592.06/5.242.18/4.142.37/4.88
102.30/4.602.31/4.363.02/6.542.30/4.612.29/4.392.66/5.31
152.86/5.382.39/4.573.25/7.482.63/5.352.35/4.402.84/5.71
253.30/6.322.51/4.663.27/7.573.31/6.232.53/4.803.09/6.31
microPET Focus 120, 350–650 keV, Fourier rebinning + 2D FBP51.92/3.661.66/3.061.90/3.811.92/3.631.65/3.091.62/3.28
101.88/3.951.74/3.221.94/3.911.83/3.691.76/3.281.66/3.34
151.99/4.021.72/3.111.98/4.051.94/3.751.77/3.221.69/3.41
252.53/4.841.73/3.012.05/4.342.45/4.491.80/3.201.81/3.67
microPET Focus 220, 250–750 keV, Fourier rebinning + 2D FBP51.751.801.701.701.701.73
101.681.781.731.661.791.75
151.821.711.801.881.721.78
252.071.691.842.091.741.87
502.881.771.982.821.821.92
754.081.902.163.921.902.11
Inveon, 350–625 keV, Fourier rebinning + 2D FBP51.63/3.361.62/3.152.45/5.621.66/3.321.63/3.141.97/4.20
101.80/3.841.58/2.912.40/5.511.72/3.401.64/3.182.12/4.44
152.03/4.321.56/2.782.29/5.321.87/3.691.63/3.052.17/4.72
252.49/5.171.61/2.862.09/4.672.38/4.761.65/2.972.06/4.54
ClearPET, 250–650 keV, 3D FBP51.94/3.762.00/4.173.24/6.052.18/4.051.97/3.923.18/5.91
101.85/3.472.27/5.973.19/5.971.87/3.682.14/4.863.20/5.88
152.01/3.622.43/5.533.20/5.962.05/3.842.33/5.253.19/5.83
252.55/4.282.42/5.693.21/5.972.50/4.182.43/6.593.19/5.85
Mosaic HP, 385–665 keV, 3D Fourier reprojection52.32/5.302.32/4.972.64/6.072.33/5.322.40/4.882.48/5.32
102.45/5.482.51/4.962.82/6.142.37/5.542.49/4.972.80/5.89
152.43/5.442.65/5.242.79/6.282.48/5.622.63/5.252.80/5.92
252.59/5.932.83/5.252.96/6.282.63/5.812.87/5.313.10/6.30
Argus, 250–700 keV, 2D FBP51.631.651.651.70
101.711.701.741.75
151.851.701.851.75
252.251.732.151.85
VrPET, 100–700 keV, SSRB + 2D FBP51.52/2.761.62/2.992.66/4.811.62/2.951.68/2.862.57/5.21
101.58/2.851.68/3.023.03/5.451.54/2.891.68/3.073.11/5.45
151.78/3.251.51/2.793.11/5.501.69/3.091.73/3.143.22/5.72
252.03/3.692.12/3.723.32/5.921.79/3.291.98/3.603.60/6.87
LabPET 8, 250–650 keV, SSRB + 2D FBP51.65/3.401.70/3.30*1.57/3.301.65/3.50*
101.91/3.601.82/3.67*1.92/3.401.74/3.45*
152.01/4.101.83/3.70*1.92/3.771.86/3.90*
252.56/4.651.90/4.28*2.55/4.701.93/4.30*
  • * Axial intrinsic resolution (FWHM/FWTM) = 1.4/4.3 mm.