TABLE 5

Summary of Counting-Rate Test Results for Mouse and Rat Phantoms

Mouse phantomRat phantom
SystemEnergy window (keV)Timing window (ns)Randoms correctionPeak NECR (kcps)Activity* (MBq)NECR at 3.7 MBq (kcps)Scatter fraction (%)Peak NECR (kcps)Activity* (MBq)NECR at 10 MBq (kcps)Scatter fraction (%)
microPET P4350–6506Calculated>601>17422.15.217325419.216.7
microPET R4350–6506Calculated61815637.29.316413730.522.2
microPET Focus 120350–6506Calculated89710366.55.626712950.920.3
microPET Focus 220250–7006Calculated>763>8947.37.235916251.819.3
Inveon350–6253.4Calculated1670131129.07.8592110137.817.2
ClearPET250–65012Calculated731829.331.0
Mosaic HP385–6657Measured5559259.65.42448765.212.7
Argus250–7007Calculated1175018.721.0404120.434.4
VrPET100–7003.8Calculated742211.5313423.3
LabPET 8250–65020Calculated2798223.515.6949119.429.5
LabPET 12250–65020Calculated3628138.916.01568340.529.3
  • * Activity at which peak NECR occurs.

  • Peak value not reached because of insufficient activity at start of scan.