Summary of Counting-Rate Test Results for Mouse and Rat Phantoms
Mouse phantom | Rat phantom | ||||||||||
System | Energy window (keV) | Timing window (ns) | Randoms correction | Peak NECR (kcps) | Activity* (MBq) | NECR at 3.7 MBq (kcps) | Scatter fraction (%) | Peak NECR (kcps) | Activity* (MBq) | NECR at 10 MBq (kcps) | Scatter fraction (%) |
microPET P4 | 350–650 | 6 | Calculated | >601† | >174† | 22.1 | 5.2 | 173 | 254 | 19.2 | 16.7 |
microPET R4 | 350–650 | 6 | Calculated | 618 | 156 | 37.2 | 9.3 | 164 | 137 | 30.5 | 22.2 |
microPET Focus 120 | 350–650 | 6 | Calculated | 897 | 103 | 66.5 | 5.6 | 267 | 129 | 50.9 | 20.3 |
microPET Focus 220 | 250–700 | 6 | Calculated | >763† | >89† | 47.3 | 7.2 | 359 | 162 | 51.8 | 19.3 |
Inveon | 350–625 | 3.4 | Calculated | 1670 | 131 | 129.0 | 7.8 | 592 | 110 | 137.8 | 17.2 |
ClearPET | 250–650 | 12 | Calculated | 73 | 18 | 29.3 | 31.0 | ||||
Mosaic HP | 385–665 | 7 | Measured | 555 | 92 | 59.6 | 5.4 | 244 | 87 | 65.2 | 12.7 |
Argus | 250–700 | 7 | Calculated | 117 | 50 | 18.7 | 21.0 | 40 | 41 | 20.4 | 34.4 |
VrPET | 100–700 | 3.8 | Calculated | 74 | 22 | 11.5 | 31 | 34 | 23.3 | ||
LabPET 8 | 250–650 | 20 | Calculated | 279 | 82 | 23.5 | 15.6 | 94 | 91 | 19.4 | 29.5 |
LabPET 12 | 250–650 | 20 | Calculated | 362 | 81 | 38.9 | 16.0 | 156 | 83 | 40.5 | 29.3 |