TABLE 4

Sensitivity over Central 7 cm of Axial FOV (Mouse Sensitivity), Complete Axial FOV (Total Sensitivity), and Peak Detection Efficiency for Each System Tested

SystemEnergy window (keV)Axial length (cm)Mouse sensitivity (%)Total sensitivity (%)Peak detection efficiency (%)
microPET P4350–6507.80.670.611.19
microPET R4350–6507.81.191.102.06
microPET Focus 220350–6507.61.261.182.28
microPET Focus 120350–6507.61.981.823.42
Inveon350–62512.74.02.86.72
ClearPET250–65011.02.321.873.03
Mosaic HP385–66511.92.431.772.83
Argus250–7004.84.32
VrPET100–7004.561.091.092.22
LabPET 8250–6507.51.451.422.36
LabPET 12250–65011.253.62.745.4