TABLE 2

Sensitivity, Specificity, and Accuracy of Lesion-Based Analysis According to Anatomic Location and Submodality

CT phase
LocationCombined CTArterial CTPortal-venous CTVenous CTPET
All (n = 354)
 Sensitivity77.1%53.4%66.1%66.9%72.8%
 Specificity85.3%92.9%92.3%89.7%97.4%
 Accuracy79.6%65.5%74.1%73.9%80.4%
Liver (n = 242)
 Sensitivity84.7%53.7%72.3%70.7%78.5%
 Specificity78.9%87.3%90.1%88.7%98.6%
 Accuracy73.8%61.3%76.4%74.8%83.1%
Lymph node (n = 58)
 Sensitivity56.9%48.3%48.3%55.2%67.2%
 Specificity85.7%95.9%89.8%85.7%100.0%
 Accuracy70.1%70.1%67.3%69.2%82.2%
Bone* (n = 23)
 Sensitivity69.6%69.6%69.6%69.6%21.7%
 Specificity100.0%100.0%100.0%100.0%100.0%
 Accuracy77.4%77.4%77.4%77.4%41.9%
Pancreas* (n = 17)
 Sensitivity58.8%52.9%52.9%52.9%94.1%
 Specificity100.0%100.0%100.0%00.0%80.0%
 Accuracy78.1%75.0%75.0%75.0%87.5%
Gastrointestinal tract* (n = 2)
 Sensitivity0.0%0.0%0.0%0.0%100.0%
 Specificity100.0%100.0%00.0%100.0%100.0%
 Accuracy50.0%50.0%50.0%50.0%100.0%
Basal lung* (n = 6)
 Sensitivity100.0%50.0%50.0%100.0%16.7%
 Specificity66.7%100.0%100.0%66.7%100.0%
 Accuracy88.9%66.7%66.7%88.9%44.4%
Other* (n = 6)
 Sensitivity50.0%50.0%50.0%50.0%83.3%
 Specificity100.0%100.0%100.0%100.0%100.0%
 Accuracy66.7%66.7%66.7%66.7%78.6%
  • * Low lesion count; cautious interpretation recommended.