TABLE 1.

Statistical Analysis of Calibration Values for 2 Scanners, With and Without Outliers Due to Known Operator Error

ScannerData setnMeanCoefficient of variation (%)95% confidence interval range for coefficient of variationMaximum change (%)
118F 2D—all data290.0546.14.8–8.220.3
18F 2D—outliers removed270.0534.23.3–5.711.2
68Ge/68Ga 2D—all data170.0493.22.4–4.99.5
18F 3D—all data280.2016.24.9–8.422.4
18F 3D—outliers removed250.1973.02.3–4.19.8
68Ge/68Ga 3D—all data130.1833.72.6–6.112.3
218F 2D—all data100.05011.07.6–20.125.7
18F 2D—outliers removed80.0534.32.9–8.88.7
68Ge/68Ga 2D—all data60.0513.11.9–7.65.0
18F 3D—all data90.1945.52.0–6.213.3
18F 3D—outliers removed80.1973.02.0–6.29.1
68Ge/68Ga 3D—all data40.1872.11.3–5.34.7
  • Units are ([scanner arbitrary units] × s/kBq × mL).