TABLE 1

Renogram Parameters for Whole-Kidney and Cortical Regions of Interest of 99mTc(CO)3(NTA) Compared with 131I-OIH in Humans (n = 9)

Left kidneyRight kidneyLeft cortexRight cortex
AgentUptake (%)TTP (min)20 min/maxUptake (%)TTP (min)20 min/maxTTP (min)20 min/maxTTP (min)20 min/max
99mTc(CO)3(NTA)49.8 (3.4)3.3 (1.6)0.15 (0.04)50.2 (3.4)3.4 (1.9)0.16 (0.04)2.52 (0.51)0.13 (0.03)2.6 (0.6)0.13 (0.02)
131I-OIH49.0 (4.1)2.6 (0.6)0.11 (0.03)51.0 (4.1)2.8 (1.6)0.11 (0.19)2.46 (0.73)0.1 (0.04)1.98 (0.6)0.10 (0.02)
P0.170.130.010.170.020.0040.710.030.020.046
  • Data are mean, with SD in parentheses.