TABLE 4

Acquisition Scheme for Pediatric CT

Tube current (mA)
WeightChestAbdomen or pelvis
4.5–8.9 (10–19)4060
9.0–17.9 (20–39)5070
18.0–26.9 (40–59)6080
27.0–35.9 (60–79)70100
36.0–45.0 (80–99)80120
45.1–70.0 (100–150)100–120140–150
>70 (>150)≥140≥170
  • Weight data are represented in kilograms, with pounds in parentheses. This table summarizes example of weight-based acquisition scheme for pediatric CT. Tube voltage for these acquisitions is 120 kVp (22).