TABLE 4

NECR Values from Patient Scans

Breast no.*
Parameter1A1U2A2U3A4A4U
Normalized injection activity (MBq)196157324270382118145
Singles (kcps)296207492412536195232
NECR (k = 1) (cps)1,474793154145107400503
NECR (k = 2) (cps)1,247668948961323401
Scatter fraction (%)30211510172525
Breast volume in PET FOV (cm3)1,0775322041853919651,001
  • * Patient number, followed by breast scanned: affected (A) or unaffected (U).

  • Dead-time–corrected, energy-windowed (350–650 keV), and energy-averaged over projections and detector heads.

  • Detector separation was 20.6 cm for patients 1 and 2 and 26.3 cm for patients 3 and 4.