TABLE 1

CT System Characteristics

ComponentCharacteristic
Radiograph tubeFocal spot: water-cooled W anode, beryllium window, 0.4 × 0.4 mm, and 0.3-mm copper-added filtration
1,000 W (i.e., maximum amperage at 80 kVp = 12.5 mA)
DetectorMaterial: indirect detection (CsI scintillator) thin film transistor
Active area: 40 × 30 cm
Native pixel matrix: 2,048 × 1,536
Native pixel size: 0.194 × 0.194 mm
2 × 2 pixel binned matrix: 1,024 × 768
Frame rate: 30 frames/s at 2 × 2 binning
  • Data are adapted from Boone et al. (37).