TABLE 1

Problems Identified During Scanner Qualification in First and Second Intervals

Frequency of problem
ProblemSubmission date, June 2005–December 2006 (n = 53)Submission date, May 2007–June 2008 (n = 48)
Overanonymization of DICOM header0 (0)4 (0)
Incorrect information in DICOM header/on application15 (1)10 (0)
Clock synchronization problem6 (0)4 (0)
Incomplete application2 (1)3 (0)
Improper data format7 (1)1 (0)
Image display problem0 (0)1 (0)
Normalization calibration5 (1)1 (1)
SUV calibration1 (0)5 (3)
Total36 (4)29 (4)
  • Data in parentheses are number of scanners that failed to pass.