TABLE 1

Acquisition and Processing Parameters for WB PET/CT and High-Resolution, Contrast-Enhanced HN PET/CT (Siemens Biograph PET/CT)

ParameterWB PET/CTHN PET/CT
Technique CTkVp, 120; mA, 240kVp, 120; mA, 200
CT collimation (mm)1.50.75
Bed time (min)612
Matrix (PET image)168256, with 1.5 zoom
Pixel size (mm)4.161.82
Filter FWHM (mm)5 (7 if >78.75 kg)2.0
Iteration46
Subset814
Intravenous contrastNoYes
Interval from injection (min)90150
BedCradleFlat
Head positionIn head holderIn head holder (chin up)
  • FWHM = full width at half maximum.