TABLE 1

Average (Mean), Maximum, and Minimum Activity Concentration and %SD in Uniform Region of Emission Images

Emission-corrected for attenuationEmission-corrected for attenuation and scatter
Transmission modeMean (kBq/mL)Maximum (kBq/mL)Minimum (kBq/mL)%SDMean (kBq/mL)Maximum (kBq/mL)Minimum (kBq/mL)%SD
CoS 1 h184.71224.58144.866.09180.13219.55140.376.24
GeS 1 h168.58203.49130.296.32164.63199.14126.456.46
GeCoi 1 h173.33230.65107.787.25168.33224.43104.307.36
GeCoiWin 1 h204.65268.01113.888.06178.05230.6698.9388.80
  • Measured phantom activity concentration: 181.5 kBq/mL.