TABLE 2

Results of Phantom Study: Percentage of Maximum Radioactivity Relative to That at 120-Second Acquisition Time (n = 4)

Acquisition time (s)
Sphere size (mm)Parameter101520253060
10Mean50.1175.0680.6787.4289.9896.97
SD9.048.622.972.722.135.10
13Mean75.0291.5998.0897.8198.9896.75
SD3.753.232.611.423.932.07
17Mean88.9895.9797.1099.2898.73100.20
SD4.011.311.830.661.751.62
22Mean96.8197.7999.4095.6996.0198.76
SD1.793.513.402.172.251.67
28Mean97.31100.4299.34100.35100.43101.86
SD2.651.321.852.032.822.15
37Mean104.4399.99100.99100.86102.62100.86
SD3.212.931.362.503.260.68