TABLE 1

Pediatric Whole-Body PET/CT Protocols Highlighting Acquisition Settings Varied for Each Weight Category

PatientCT acquisition, maximum tube current for CT (mAs)PET acquisition
CategoryWeight range (kg)Estimated whole-body scan range (mm)Injected activity (MBq [mCi])Scan time per FOV (min)
Pink6–7.459.5–66.51037 (1.0)3
Red7.5–9.466.5–741045 (1.2)3
Purple9.5–11.474–84.51555 (1.5)3
Yellow11.5–14.484.5–97.52068 (1.9)3
White14.5–18.497.5–1102087 (2.4)3
Blue18.5–22.4110–12220108 (2.9)3
Orange22.5–31.4122–13725142 (3.9)5
Green31.5–40.5137–15030190 (5.1)5
Black40.5–55150–16230252 (6.8)5
Small adult55–70162–17035330 (8.9)5
Large adult>70>17040370 (10.0)5