Sphere size (mm) | |||||||
---|---|---|---|---|---|---|---|
Acquisition time | Parameter | 10 | 13 | 17 | 22 | 28 | 37 |
15 min | 2,881.29 | 4,896.79 | 5,548.12 | 6,111.37 | 6,370.27 | 6,500.82 | |
3 min | Mean (Bq/mL) | † | 4,670 | 5,504 | 6,297 | 6,640 | 7,352 |
SD | 155 | 257 | 448 | 174 | 273 | ||
CV | 3.32 | 4.66 | 7.12 | 2.62 | 3.71 | ||
1 min | Mean (Bq/mL) | † | 4,517 | 5,547 | 6,602 | 7,282 | 8,240 |
SD | 428 | 211 | 214 | 704 | 530 | ||
CV | 9.48 | 3.81 | 3.25 | 9.67 | 6.44 | ||
%Dif-index* | −3.29 | 0.78 | 4.85 | 9.67 | 12.07 | ||
30 s | Mean (Bq/mL) | † | 2,501 | 4,986 | 6,792 | 8,486 | 9,469 |
SD | 745 | 761 | 683 | 1,232 | 426 | ||
CV | 29.78 | 15.26 | 10.06 | 14.52 | 4.50 | ||
%Dif-index* | −46.45 | −9.43 | 7.86 | 27.80 | 28.79 | ||
15 s | Mean (Bq/mL) | † | † | 4,384 | 5,544 | 7,327 | 9,275 |
SD | 630 | 709 | 1,208 | 868 | |||
CV | 14.36 | 12.79 | 16.49 | 9.36 | |||
%Dif-index* | −20.35 | −11.96 | 10.34 | 26.15 |