TABLE 4

Results of Phantom Study for an 18F-FDG Concentration of 7 kBq/mL (n = 5)

Sphere size (mm)
Acquisition timeParameter101317222837
15 min2,881.294,896.795,548.126,111.376,370.276,500.82
3 minMean (Bq/mL)4,6705,5046,2976,6407,352
SD155257448174273
CV3.324.667.122.623.71
1 minMean (Bq/mL)4,5175,5476,6027,2828,240
SD428211214704530
CV9.483.813.259.676.44
%Dif-index*−3.290.784.859.6712.07
30 sMean (Bq/mL)2,5014,9866,7928,4869,469
SD7457616831,232426
CV29.7815.2610.0614.524.50
%Dif-index*−46.45−9.437.8627.8028.79
15 sMean (Bq/mL)4,3845,5447,3279,275
SD6307091,208868
CV14.3612.7916.499.36
%Dif-index*−20.35−11.9610.3426.15
  • * %Dif-index = percentage difference compared with 3-min acquisition.

  • Difficult to distinguish from background noise.