TABLE 2

Results of Phantom Study for an 18F-FDG Concentration of 53 kBq /mL (n = 5)

Sphere size (mm)
Acquisition timeParameter101317222837
15 min25,40540,09850,27552,46653,63853,692
3 minMean (Bq/mL)24,69040,02950,16952,65754,68754,788
SD777804130110481254627
CV3.152.012.591.992.291.14
1 minMean (Bq/mL)23,59138,98150,51552,67455,62156,965
SD1,3392,2072,064687741340
CV5.685.664.091.301.330.60
%Dif-index*−4.45−2.620.690.031.713.97
30 sMean (Bq/mL)20,85939,07650,33154,19657,46259,467
SD1,4222,3802,1651,2961,1561,264
CV6.826.094.302.392.012.13
%Dif-index*−15.52−2.380.322.925.078.54
15 sMean (Bq/mL)16,34637,66850,07855,98958,99362,000
SD2,9003,2104,1672,9303,4512,054
CV17.748.528.325.235.853.31
%Dif-index*−33.79−5.90−0.186.337.8713.16