TABLE 2

Average Crystal Energy (FWHM) and DOI Resolutions (FWHM) Measured at 3 Different Temperatures

Temperature (°C)
Resolution01020
Energy (%)14.314.616.2
DOI (mm)3.163.163.68
  • Lower energy threshold was 350 keV. Radiation beam width was 2 mm for DOI measurements.