Summary of Recommendations for Routine QC of Clinical Nuclear Medicine Instrumentation
Instrument | At installation or after repair | Annual | Quarterly or monthly | Weekly | Daily |
---|---|---|---|---|---|
Survey meter | Accuracy | Accuracy | —* | —* | Battery check |
Background | |||||
Constancy | |||||
Dose calibrator | Geometry (volume)- dependent response | N/A | Linearity | —* | Constancy |
Accuracy | |||||
Well counter | Efficiency (sensitivity) | Efficiency (sensitivity) | —* | —* | Energy peaking Background |
Constancy | |||||
Intraoperative probe | —* | —* | —* | —* | Battery check |
Bias check | |||||
Energy peaking (if applicable) | |||||
Background | |||||
Constancy | |||||
Organ uptake probe | —* | —* | —* | —* | Energy peaking |
Background | |||||
Constancy | |||||
Efficiency (sensitivity) | |||||
γ-camera | References 3, 5, and 8 | —* | Uniformity (for radionuclides used clinically other than 99mTc or 57Co) | Spatial resolution | Energy peaking Uniformity (for 99mTc or 57Co) |
SPECT scanner or SPECT subsystem of SPECT/CT scanner | Reference 6 | —* | Tomographic uniformity | COR alignment | Energy peaking |
Overall system performance | Uniformity (for 99mTc or 57Co) | ||||
PET scanner or PET subsystem of PET/CT scanner | Reference 4 | Normalization | —* | Tomographic uniformity | Blank scan |
Normalization | Well-counter calibration | ||||
Well-counter calibration | |||||
CT subsystem of PET/CT or SPECT/CT scanner | References 1, 2, and 22 | Dosimetry | Laser alignment | —* | Warm-up |
Image slice thickness | Tomographic uniformity | ||||
Spatial resolutionLinearity | Accuracy of CT number of water | ||||
High-contrast resolution | |||||
Low-contrast resolution | Noise | ||||
PET/CT or SPECT/CT image registration | —* | —* | Accuracy | —* | —* |
CT-based attenuation corrections |
↵* Other than QC or performance assessment procedures that may be recommended by manufacturer, no such procedure specifically recommended for this time interval exists.