TABLE 2

Radiation Dose and Dose Equivalent from different Protocols for Acquiring CT Attenuation Data

RestStressTotal
CT protocolmGymSvmGymSvmGymSvTotal CT + Rb (mSv)
Helical CT* rest + stress5.71.335.71.3311.42.667.10
Cine rest + stress + obligatory helical CT102.36102.3622.95.399.83
2.90.67
Cine stress + low-dose obligatory helical CT2.90.67102.3612.93.037.47
82Rb 1,850 MBq, intravenous2.222.224.44
  • * Helical CT (200 mA) used for attenuation correction of emission data.

  • Obligatory helical CT (100 mA) required by commercial software to complete image acquisition but not used for attenuation correction.

  • + = addition to or cumulative dose.