TABLE 1

Image Resolution of VP-PET Systems Using Detectors of Different Sizes

Detector 1
12 × 12 LSO crystals8 × 8 LSO crystals
1.51 × 1.51 × 10.0 mm32.2 × 2.2 × 10.0 mm3
13 × 13 LSO crystals(a) 2.40 ± 0.06 mm(b) 2.76 ± 0.04 mm
4.0 × 4.0 × 20.0 mm3(2.30 mm)(2.72 mm)
Detector 28 × 8 LSO crystals(c) 3.00 ± 0.02 mm(d) 3.24 ± 0.12 mm
6.4 × 6.4 × 25.0 mm3(2.83 mm)(2.83 mm)
  • Average resolution (FWHM) was measured from 4 point sources imaged by different combinations of detectors. Predicted image resolution of 4 systems was calculated by Equation 8 and is shown in parentheses.