TABLE 4

Patients with Cine CT PET Having Worse Misregistration Artifactual Defects than with Helical CT PET Corrected by Shifting Cine CT Data

Cine CT PET worse than helical CT (8 patients)Severity (minimum quadrant average)Size–severity (% <60% maximum)Size (% ≤2.5 SD)Transaxial misregistration (mm)
Helical CT PET74.79 ± 3.601.72 ± 2.585.26 ± 4.498.39 ± 3.00
Cine CT PET71.28 ± 5.459.64 ± 10.2710.08 ± 5.5712.44 ± 5.51
n = 8, P for Δ0.02160.02840.00180.0751
Cine CT PET70.29 ± 5.0711.01 ± 10.2710.88 ± 5.4912.24 ± 5.93
Shifted cine CT PET75.98 ± 6.104.05 ± 6.604.98 ± 5.772.71 ± 3.39
n = 7*, P for Δ0.00430.06390.00470.0008
  • * Quantitative shifted cine CT data were lost for 1 patient due to corruption of the header but with good saved images.