TABLE 2

CRC and Variability Numbers from NEMA/IEC Torso Phantom Reconstructions for 3-Minute Scan Time

Sphere diameter (mm)
Reconstruction algorithmIteration no.Parameter101317222837
UPenn TOF6CRC (%)354770697482
Variability (%)6.96.15.24.43.93.2
UPenn non-TOF6CRC (%)314269707182
Variability (%)5.75.24.63.93.12.5
DLT3CRC (%)354466655970
Variability (%)4.84.64.13.73.33.0
LOR RAMLA2CRC (%)233460626878
Variability (%)6.35.95.14.43.62.9