TABLE 2

Correlation Between Manual 2D ROI, MRI-Based, and Automated 3D VOI Semiquantitative Evaluations

Evaluation
ParameterManual vs. automated (n = 80)Manual vs. MRI-based (n = 40)Automated vs. MRI-based (n = 40)
R20.9340.8420.871
Constant ± SE−0.120 ± 0.038−0.122 ± 0.1070.083 ± 0.083
Slope ± SE1.115 ± 0.0341.228 ± 0.0861.040 ± 0.065
Limits of agreement−0.12 to 0.12−0.19 to 0.19−0.18 to 0.18
ICC0.9660.9160.932
  • R2, constant, and slope refer to a linear regression line. Limits of agreement (according to Bland and Altman (10)) and ICCs were calculated after correcting data for different slopes.