Relative distribution (%) | |||||||
---|---|---|---|---|---|---|---|
Volume | Step | Eluent (concentration) | Ge(III) | Ge(IV) | Zn(II) | Ti(IV) | Fe(III) |
Set 1 | |||||||
7 mL | Generator elution | 0.10N HCl | 0.19 | 98.14 | 0.61 | 5.34 | 0.45 |
5 mL | Purification | 80% acetone/0.10N HCl | 0.58 | 1.83 | 99.39 | 2.68 | 53.75 |
0.4 mL | Ga(III) elution | 97.6% acetone/0.05N HCl | 98.50 | 2·10−2 | 5·10−2 | 5·10−2 | 43.54 |
1 mL | Washing | 4N HCl | 0.53 | 6·10−3 | <10−3 | 90.92 | 2.01 |
1 mL | Washing | H2O | 0.20 | 5·10−3 | <10−3 | 1.01 | 0.29 |
Set 2 | |||||||
7 mL | Generator elution | 0.10N HCl | 0.16 | 97.08 | 0.77 | 7.30 | 0.13 |
0.6 mL | Purification | 80% acetone/0.15N HCl | 1.43 | 2.92 | 98.15 | 0.68 | 37.86 |
0.4 mL | Ga(III) elution | 97.6% acetone/0.05N HCl | 97.82 | 3·10−2 | 1.08 | 7·10−2 | 49.78 |
1 mL | Washing | 4N HCl | 0.41 | 5·10−3 | 5·10−3 | 72.15 | 11.61 |
1 mL | Washing | H2O | 0.18 | 3·10−3 | <10−3 | 19.80 | 0.62 |
Set 3 | |||||||
7 mL | Generator elution | 0.10N HCl | 0.11 | 97.68 | 0.43 | 7.18 | 0.73 |
5 mL | Purification | 80% acetone/0.15N HCl | 6.29 | 2.32 | 99.57 | 3.99 | 87.37 |
0.4 mL | Ga(III) elution | 97.6% acetone/0.05N HCl | 92.73 | 5·10−3 | <10−3 | 0.11 | 11.10 |
1 mL | Washing | 4N HCl | 0.77 | 2·10−3 | <10−3 | 86.54 | 0.71 |
1 mL | Washing | H2O | 0.10 | 2·10−3 | <10−3 | 2.18 | 0.09 |
Set 4 | |||||||
7 mL | Generator elution | 0.10N HCl | 0.29 | 98.08 | 0.92 | 6.33 | 0.4 |
0.6 mL | Purification | 80% acetone/0.20N HCl | 6.34 | 1.86 | 96.58 | 1.11 | 54.0 |
0.4 mL | Ga(III) elution | 97.6% acetone/0.05N HCl | 92.96 | 5·10−2 | 3.70 | 0.11 | 42.6 |
1 mL | Washing | 4N HCl | 0.28 | 6·10−3 | <10−3 | 87.37 | 2.9 |
1 mL | Washing | H2O | 0.13 | 6·10−3 | <10−3 | 5.08 | 0.1 |