TABLE 1

Relative Distribution of Ga(III), Ge(IV), Zn(II), Ti(IV), and Fe(III) on Microchromatographic Column (50 mg of Bio-Rad AG 50W-X8 resin, 400 Mesh) in Hydrochloric Acid/Acetone Media

Relative distribution (%)
VolumeStepEluent (concentration)Ge(III)Ge(IV)Zn(II)Ti(IV)Fe(III)
Set 1
 7 mLGenerator elution0.10N HCl0.1998.140.615.340.45
 5 mLPurification80% acetone/0.10N HCl0.581.8399.392.6853.75
 0.4 mLGa(III) elution97.6% acetone/0.05N HCl98.502·10−25·10−25·10−243.54
 1 mLWashing4N HCl0.536·10−3<10−390.922.01
 1 mLWashingH2O0.205·10−3<10−31.010.29
Set 2
 7 mLGenerator elution0.10N HCl0.1697.080.777.300.13
 0.6 mLPurification80% acetone/0.15N HCl1.432.9298.150.6837.86
 0.4 mLGa(III) elution97.6% acetone/0.05N HCl97.823·10−21.087·10−249.78
 1 mLWashing4N HCl0.415·10−35·10−372.1511.61
 1 mLWashingH2O0.183·10−3<10−319.800.62
Set 3
 7 mLGenerator elution0.10N HCl0.1197.680.437.180.73
 5 mLPurification80% acetone/0.15N HCl6.292.3299.573.9987.37
 0.4 mLGa(III) elution97.6% acetone/0.05N HCl92.735·10−3<10−30.1111.10
 1 mLWashing4N HCl0.772·10−3<10−386.540.71
 1 mLWashingH2O0.102·10−3<10−32.180.09
Set 4
 7 mLGenerator elution0.10N HCl0.2998.080.926.330.4
 0.6 mLPurification80% acetone/0.20N HCl6.341.8696.581.1154.0
 0.4 mLGa(III) elution97.6% acetone/0.05N HCl92.965·10−23.700.1142.6
 1 mLWashing4N HCl0.286·10−3<10−387.372.9
 1 mLWashingH2O0.136·10−3<10−35.080.1