TABLE 3

Percent Contrast, Background Variability, and Average Lung Residual of SET-3000 G/X Scanner Measured with Image Quality Phantom

Sphere diameter (mm)Lung insert (30 mm)
Parameter10.013.017.022.028.037.0
L/B = 4
Hot-sphere recovery coefficient (%)17.826.338.965.8
Cold-sphere contrast (%)53.261.3
Background variability (%)11.210.910.59.37.56.9
Average residual (%) over lung insert30.2
L/B = 8
Hot-sphere recovery coefficient (%)18.527.739.066.1
Cold-sphere contrast (%)50.056.8
Background variability (%)9.79.68.46.96.05.7
Average residual (%) over lung insert29.3
  • L/B = ratio of lesion activity to background activity.