TABLE 6

SPECT-Based DPM Results of Mean Absorbed Dose for Targets in Voxel Phantom with HE Collimator and UHE Collimator

TargetTrue dose (cGy)SPECT-derived mean dose (cGy)Relative SD (%)Error (%)
HE collimator
7-mL tumor1,3098517.235
16-mL tumor1,3451,2643.06
59-mL tumor1,4501,4311.91
135-mL tumor1,4581,5120.9−4
Liver5005600.4−12
Kidney1,1601,1280.63
Spleen8138590.7−6
UHE collimator
7-mL tumor1,3098987.931
16-mL tumor1,3451,2741.95
59-mL tumor1,4501,3571.06
135-mL tumor1,4581,4930.8−2
Liver5005020.30
Kidney1,1601,0990.45
Spleen8138130.60