TABLE 4

SPECT Activity Quantification Accuracy (% Error) with Different Calibration Geometries

TargetPoint source calibrationHot-sphere calibrationUniform tank calibration
VOI radius = 2 pixelsVOI radius = 4 pixels
HEUHEHEUHEHEUHEHEUHE
7-mL tumor34.336.541.542.440.838.547.740.7
16-mL tumor−0.35.110.713.99.68.020.111.4
59-mL tumor−3.47.08.015.66.99.917.613.2
135-mL tumor−11.7−4.10.55.5−0.6−0.911.02.8
Liver−20.8−1.1−7.58.3−8.82.03.85.6
Kidney−1.86.29.314.88.39.118.912.4
Spleen−12.30.10.09.3−1.23.210.56.7